Radio FrequencyForum -
related threads
 
(RF) Radio frequency refers to that portion of the electromagnetic spectrum in which electromagnetic waves can be generated by alternating current fed to an antenna.
The RF pulses used in MRI are commonly in the 1-100 megahertz range, and their effect upon a body is potential heating of tissues and foreign bodies, such as metallic implants, mainly at the surface.
This is a principal area of concern for MRI safety limits caused by absorption (see specific absorption rate) of the applied pulses of RF energy.
mri safety guidance
Image Guidance
The higher the frequency, the larger will be the amount of heat developed.
The more ionic the biochemical environment in the tissue, the more energy that will be deposited as heat.
This effect is well known for homogeneous model systems, but the complex structure of various human tissues makes detailed theoretical calculations very difficult, if not impossible. By scanning problems, it is important to verify the transmission frequency. If the RF transmitted into the patient was, e.g. 5000 Hz lower than the resonance frequency of the protons, no protons was excited, and no signal returns.
spacer
 
• Related Searches:
    • Adiabatic Radio Frequency Pulses
    • Magnetic Field
    • Device
    • Absorbed Dose
    • Larmor Frequency
 
Further Reading:
  Basics:
Radio frequency
   by en.wikipedia.org    
  News & More:
On the estimation of the worst-case implant-induced RF-heating in multi-channel MRI.
Thursday, 2 March 2017   by www.ncbi.nlm.nih.gov    
Safety and reliability of Radio Frequency Identification Devices in Magnetic Resonance Imaging and Computed Tomography
Thursday, 4 February 2010   by 7thspace.com    
Radiologists Offer Non-surgical Treatment For Early-stage Liver Cancer
Saturday, 5 March 2005   by www.sciencedaily.com    
MRI Resources 
Collections - Most Wanted - Education pool - Bioinformatics - Directories - MRI Training Courses
 
Radio Frequency Shielding
 
Radio frequency shielding includes the construction of enclosures for the purpose of reducing the transmission of electric or magnetic fields from one space to another (Faraday cage, Faraday shield). Electrically conducted shielding is designed to isolate MRI systems from its environment at the resonant frequencies.
All electronic and computer systems radiate certain frequencies of radio and magnetic waves. They can interfere with other equipment in the vicinity. Magnetic shielding enclosures are used to reduce the levels of RF radiation that enters or leaves the shielded room.
Copper shielding enclosures are designed to filter a range of frequencies under specified conditions. One of the characteristics of copper is its high electrical conductivity. Also its other physical properties like ductility, malleability, and ease of soldering, make it an ideal material for radio frequency shielding. Sheet copper can be formed into any shape and size, and electrically connected to a grounding system to provide an effective RF shielding.

See also MRI Safety
spacer

• View the DATABASE results for 'Radio Frequency Shielding' (3).Open this link in a new window

MRI Resources 
Cardiovascular Imaging - Pathology - MRI Centers - Case Studies - Intraoperative MRI - Blood Flow Imaging
 
Skin Depth
 
Time-dependent electromagnetic fields are significantly attenuated by conducting media (including the human body); the skin depth gives a measure of the average depth of penetration of the RF field. A high power frequency tunable RF source can be rapidly switched on and off. This produces a large RF field perpendicular to the magnetic field. This RF field is focused by the body coil. The RF source and coils must be tunable in both frequency and impedance to 'match the impedance' of the patient's body.
mri safety guidance
MRI Safety Guidance
The skin depth may be a limiting factor in MR imaging at very high frequencies (high magnetic fields). The skin depth also affects the Q of the coils.
spacer
 
Further Reading:
  News & More:
Magnetic resonance-guided motorized transcranial ultrasound system for blood-brain barrier permeabilization along arbitrary trajectories in rodents
Thursday, 24 December 2015   by www.ncbi.nlm.nih.gov    
Searchterm 'MRI Safety' was also found in the following services: 
spacer
News  (32)  Resources  (30)  Forum  (2)  
 
Skin PatchesMRI Resource Directory:
 - Safety -
 
Pharmaceutical products in transdermal skin patches may cause burns due to the absorption of RF energy.
mri safety guidance
MRI Safety Guidance
Such patches must be removed prior to MR examinations.
spacer
MRI Resources 
Safety Training - Veterinary MRI - IR - Used and Refurbished MRI Equipment - Examinations - Chemistry
 
TattoosMRI Resource Directory:
 - Safety -
 
Tattoos with metallic or ferromagnetic color ingredients can distort MR images. Skin irritations or burns caused by heating of the metallic particles due to the radio frequency pulse are also possible, but rare.
mri safety guidance
MRI Safety Guidance
Patients should be informed about the possible risk, and scanning should be stopped immediately if they feel heat. Metallic make-up can even be pulled into the eye by magnetic forces and should be removed previous to the MRI scan.
spacer

• View the DATABASE results for 'Tattoos' (2).Open this link in a new window

 
Further Reading:
  News & More:
Going for an MRI scan with tattoos? First prospective study on risk assessment
Wednesday, 30 January 2019   by www.cbs.mpg.de    
Women with permanent make-up tattoos suffer horrific facial burns after going in for MRI scans - which create an electric current in the ink
Monday, 4 July 2016   by www.dailymail.co.uk    
MRI Resources 
Safety Training - PACS - Case Studies - Claustrophobia - MR Myelography - MRI Technician and Technologist Jobs
 
previous      36 - 40 (of 42)     next
Result Pages : [1]  [2 3 4 5 6 7 8 9]
 Random Page
 
Share This Page
FacebookTwitterLinkedIn

MR-TIP    
Community   
User
Pass
Forgot your UserID/Password ?    



New acceleration techniques will :
reduce scan times 
cause artifacts 
increase expenses 
be useful if you have a lot of experience 
doesn't do much 
never heard of 

Look
      Ups





MR-TIP.com uses cookies! By browsing MR-TIP.com, you agree to our use of cookies.

Magnetic Resonance - Technology Information Portal
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology-TIP • Medical-Ultrasound-Imaging • 
Copyright © 2003 - 2024 SoftWays. All rights reserved. [ 2 May 2024]
Terms of Use | Privacy Policy | Advertising
 [last update: 2024-02-26 03:41:00]